Yes. We support OEM/ODM, custom logo, packaging and in-depth function customization are available.
-
Structural parts: 2 years; wearable components (e.g., tires): 6 months. Extended warranty available upon request.
-
Use reinforced wooden crates with shockproof materials. International logistics standards (e.g., ISTA 3A) must be followed.
-
Follow ISO 7176 series (e.g., ISO 7176-8 for static load testing). Fatigue tests simulate 200,000 cycles of use.
-
CE certification under MDR (EU 2017/745) requires ISO 13485 quality system + clinical evaluation. Class I devices can self-declare, while Class II/III need Notified Body approval.
-
Seat depth should support 80% thigh length. Pressure-relief cushions comply with ISO 16840-2.
-
Compliance with EN 11199-1 for stability and ISO 24415-1 for dry surface friction coefficient (≥0.3).
-
Static load testing per ISO 7176-8 (e.g., bed frame load ≥200kg). Durability tests simulate ≥10,000 adjustment cycles.
-
Hospital beds must pass static load testing ≥200kg according to ISO 7176-8 . For dynamic durability, motors must withstand ≥10,000 cycles under full load.
-
Headrest tilt ≥65°, knee tilt ≥25°, and full-bed tilt ≥12° must be tested for accuracy and stability under YY9706.252-2021.
-
Guardrails must be tested for lateral/vertical forces (100N, 3,000 cycles) and withstand static loads without unlocking . Height ≥0.35m, gap <0.12m .
-
Materials in contact with skin must comply with ISO 10993-1 (cytotoxicity, sensitization) . Metal surfaces require anti-corrosion treatment .
-
Tests include insulation resistance (≥10MΩ),leakage current (≤0.1mA),and grounding resistance (≤0.2Ω) per IEC 60601-1 .